| SPECIFICATIONS |
| MODEL |
7130 |
7132 |
7140 |
7142 |
| AC WITHSTAND VOLTAGE |
| Output Rating |
5KVAC/20mA |
Range |
Resolution |
Accuracy |
| Output Voltage, KVAC |
0-5.00 |
0.01 |
±(2% of setting + 5V) |
| Output Frequency |
50Hz/60Hz ±100ppm, User Selection |
| Output Waveform |
Sine Wave ,THD.<2% (Resistive Load), Crest Factor = 1.3 - 1.5 |
| Output Regulation |
± (1% of output + 5V), From no load to full load |
| SETTINGS |
| Max-Limit AC Current, mA |
0 - 20.00 |
0.01 |
±(2% of setting + 2counts) |
| Min-Limit AC Current, mA |
0 - 9.999 |
0.001 |
±(2% of setting + 2counts) |
| Ramp Up Time, second |
0.1 - 999.9 |
0.1 |
±(0.1% + 0.05sec) |
| Ramp Down Time, second |
0 - 999.9 |
0.1 |
| Dwell Time, second |
0, 0.3 - 999.9
(0=continuous) |
0.1 |
| Arc Detection |
0, 1 - 9 ranges (0=OFF, 9 is the most sensitivity) |
| DC WITHSTAND VOLTAGE (7140,7142 ONLY) |
| Output Rating |
6KVDC/7500µA |
| Output Voltage, KVDC |
0 - 6.00 |
0.01 |
±(2% of setting + 5V) |
| Output Ripple |
<5%(6KV/7500µA at Resistive Load) |
| SETTINGS |
| Max-Limit DC Current, µA |
0 - 7500 |
1 |
±(2% of setting + 2counts) |
| Min-Limit DC Current, µA |
0 - 999.9 |
0.1 |
±(2% of setting + 2counts) |
| Ramp Up Time, second |
0.1 - 999.9 |
0.1 |
±(0.1% + 0.05sec) |
| Ramp Down Time, second |
0, 1.0 - 999.9 |
0.1 |
| Dwell Time, second |
0, 0.4 - 999.9
(0=continuous) |
0.1 |
| Arc Detection |
0, 1 - 9 ranges (0=OFF, 9 is the most sensitivity) |
| Discharge Time |
≦ 200 msec |
| INSULATION RESISTANCE (7132,7142 ONLY) |
| Output Rating |
1KVDC/9999MΩ |
| Output Voltage, DCV |
30 - 1000 |
10 |
±(2% of setting + 5V) |
| SETTINGS |
| Max-Limit Resistance, MΩ |
0, 1 - 9999 (0=OFF) |
1 |
±(2% of setting+2counts)at 500-1000VDC,
1 -1000 MΩ
|
| Min-Limit Resistance, MΩ |
1 - 9999 |
1 |
±(5% of setting+2counts)at 500-1000VDC,
1000 -9999 MΩ |
±(8% of setting+2counts)at 30-500VDC,
1 -1000 MΩ |
| Ramp Up Time, second |
0.1 - 999.9 |
0.1 |
±(0.1% + 0.05sec) |
| Ramp Down Time, second |
0, 1.0 - 999.9 |
0.1 |
| Delay Time, second |
0, 0.5 - 999.9
(0=continuous) |
0.1 |
| MEASUREMENT |
| AC/DC Voltage, KV |
0 - 6.00 |
0.01 |
±(1.5% of reading + 1count) |
| DC Voltage, V (IR only) |
30 - 1000 |
1 |
±(1.5% of reading + 5V) |
| AC Current, mA |
0 - 3.500 |
0.001 |
±(2% of reading + 2counts) |
3.00 - 20.00 |
0.01 |
| DC Current, µA |
0 - 350.0 |
0.1 |
±(2% of reading + 2counts) |
| DC Current, mA |
0.300 - 3.500 |
0.001 |
3.00 - 7.50 |
0.01 |
| Resistance, , MΩ |
1 - 9999
(Auto Range) |
0.001
0.01
0.1
1 |
±(2% of setting+2counts)at 500-1000VDC,
1 -1000 MΩ
±(5% of setting+2counts)at 500-1000VDC,
1000 -9999 MΩ
±(8% of setting+2counts)at 30-500VDC,
1 -1000 MΩ |
| GENERAL |
| Input Voltage AC |
115/230Vac±15%, 50/60Hz ± 5%, Fuse 3.15A |
| PLC Remote Control |
Input : Test , Reset , Memory1.2.3 , Interlock |
| Output : Pass , Fail , Processing , Reset-Out |
| Interface(Option) |
RS485 Interface |
| Memory |
10 memories, 3 steps/memory |
| Graphic Display |
128×64 Graphic LCD |
| Safety |
Built-in Smart GFI circuit,GFI trip current 450?A max, HV shut down speed: <1mS |
| Key Lock |
To prevent unauthorized alteration of the test Parameters |
| Calibration |
Build-in software and external calibrated meters |
| Verification |
Build-in software driven verification menu to test fault detection circuits |
| Alarm Volume Setting |
Range: 0-9 ;0=OFF, 1 is softest volume, 9 is loudest volume. |
| Environment |
0-40℃, 20-80%RH |
| Dimension/Net Weight |
215mm(W)×89mm(H)×370mm(D)/ 10Kg |
| STANDARD ACCESSORIES |
| Power Cord(10A) |
×1 |
| Fuses |
×2 (Including a spare contained in the fuse holder) |
| Interlock Disable Key(1505) |
×1 |
| Hipot Test Lead ,1.5m(1101) |
×1 |
| Hipot Return Lead ,1.8m(1102) |
×1 |
| *product specifications are subject to change without notice. |
|